 Oxygraph-2k Trouble ShootingYou must login first to activate the following links.
- Oxygen signal not responding
Low oxygen signal of the polarographic oxygen sensor (O2S) even at air saturation.
High zero signal due to polarographic oxygen sensor (POS) or electronics.
- Oxygen signal an POS connector
The signal changes when the O2S connector is touched or moved.
- Oxygen calibration changes suddenly
Check O2k setup settings in DatLab, POS membrane, or chamber assembly.
- DatLab error messages - problems and solutions.
- Flux indicates an apparently low RCR
In experiments with isolated mitochondria, state 2 - state 3 - state 4 transitions are used for calculating the P/O ratio and respiratory control ratio (RCR). Even with a good mitochondrial preparation, the RCR may appear to be low when observing on-line the peak oxygen flux at state 3. - There is a simple solution.
Why does the plot of oxygen flux apparently lag behind the time course of oxygen concentration?
Why does oxygen flux reach a low peak value (and a correspondingly low RCR is obtained) when an ADP pulse is added to isolated mitochondria?
- Lasting effect of EtOH-soluble inhibitors
Oxygen flux per mg tissue was significantly lower (<50 %) in one particular chamber, compared to parallel experiments in the other chamber(s). The effect was due to incomplete washout of inhibitors used in the titration protocol.
Failure of Temperature Control
Recorder Interface